22. Internal Pulse ccuracy (Option 002)
For B-Series Only)
Description and Procedure
his performance test uses an oscilloscope to measure the pulse width of the output of the
modulation generator (not the pulse width of the RF output).
1. Connect the equipment as shown in Figure 2-31. Preset all instruments and let them warm
up for at least one hour.
2. On the synthesizer, press:
Pulse Menu Pulse On/Off Int
MOD
5
Internal Menu Internal Pulse Width
Internal Pulse Period
Agilent 8360
Figure 2-31. Pulse Accuracy Test etup
Equipment
Digitizing Oscilloscope
Oscilloscope Probes
(asterisk on)
300
ns
5
5
HP/Agilent 54111D
HP/Agilent 10437A
25
ns
5
5
Internal Pulse Accuracy
Performance Tests
2-105