Frequency Range And Accuracy Check - Agilent Technologies N5161A Installation Manual

Mxg signal generators
Hide thumbs Also See for N5161A:
Table of Contents

Advertisement

Operation Verification

Frequency Range and Accuracy Check

Frequency Range and Accuracy Check
The frequency range is tested by determining the frequency accuracy relative to the timebase at the frequency limits of
the signal generator. This test can be performed with either a frequency counter or a spectrum analyzer.
Table 3-1 Recommended Equipment for Checking N5161A/62A/81A/82A/83A Frequency Range and Accuracy
Test Equipment
Frequency Counter
Spectrum Analyzer
E4440A
E4443A
E4445A
E4446A
E4447A
E4448A
20
N5161A/62A/81A/82A:
Agilent 53131A or 53132A with Option 010 and 050 or 124
N5161A/62A/81A/82A
Y
Y
Y
Y
Y
Y
Agilent N5161A/62A/81A/82A/83A MXG Signal Generators Installation Guide
Recommended Model
N5183A Option 520
Y
--
--
Y
Y
Y
N5183A Option 532/540
--
--
--
Y
Y
Y

Advertisement

Table of Contents
loading

This manual is also suitable for:

N5162aN5181aN5182aN5183a

Table of Contents