Agilent Technologies 8360 B Series Service Manual page 294

Swept signal/cw generator
Hide thumbs Also See for 8360 B Series:
Table of Contents

Advertisement

Procedure
1. Connect the equipment as shown in Figure 3-18 using an analog oscilloscope with A
versus B sweep capability. Do not connect the power meter yet. (See Figure 3-19 for
location detail of XA10J2 pin 55 on the motherboard.) Power on all the instruments and
let them warm up for at least one hour.
2. Press
on the synthesizer and, for the 83640B/L and 83650B/L press:
PRESET
5
Tools Menu Disable Doubler
SERVICE
5
3. On the power meter:
Zero and calibrate the power meter/sensor.
Set the power meter to dBm mode.
Analog Oscilloscope
Oscilloscope Probes
Power Meter
Power Sensor
Power Sensor
Attenuator 10 dB
Attenuator 10 dB
Agilent 8360
Figure 3-18. Amplifier/Filter Adjustments etup
Equipment
HP/Agilent 1740A
HP/Agilent 10431A
HP/Agilent 436A/7A/8A
HP/Agilent 8485A (83620B/22B/24B,
83623B/L, and 83630B/L)
HP/Agilent 8487A (83640B/L and 83650B/L)
HP/Agilent 8493C Option 010 (83620B/22B/24B,
83623B/L, and 83630B/L)
HP/Agilent 8490D Option 010 (83640B/L and 83650B/L)
(asterisk on)
Amplifier/Filter Adjustments
Adjustments
3-37