To calibrate the oscilloscope

Your Infiniium should be at optimum performance when performing mask
testing. If your Infiniium has not been calibrated in the last year or if the
Calibration ∆Temp is greater than ±5 °C, you should recalibrate the oscilloscope
before performing a mask test. You can find the Calibration ∆Temp in the
Calibration dialog box by clicking on the Utilities menu and selecting Calibration.
Use a High-Quality Cable
When calibrating the oscilloscope, use a short, high-quality BNC-to-BNC coaxial
cable, such as the 8120-1838 cable.
Make sure Infiniium has warmed up for at least 30 minutes.
1
From the oscilloscope main menu click
2
Make sure that the
3
Cal Memory Protect
This option protects the oscilloscope's calibration factors from
accidentally being changed.
Details
When checked, this option displays a list of anything that has failed
during the calibration. You can use the failure information to help repair
the oscilloscope.
Click
Start
.
4
Infiniium will display additional instructions and guide you through the step-
by-step calibration procedure.
check box is unchecked.
Cal Memory Protect
Getting Started and Using Masks
To calibrate the oscilloscope
, then
Utilities
Calibration
.
2-5