Final Steps

Customer-specific and Country-specific Settings for the Multiline

Set requested language according to instruction RXL2-120.032... "Service Software Soft-
ware parameter/monitoring". Parameter Group: / Configuration: User terminal.
Set clinic name according to instruction RXL2-120.032... "Service Software Software
parameter/monitoring. Parameter Group: / Clinic data.
NOTICE
Correct the date and time in the host computer D1 according to instruction "Operation of
the service software" RXL2-120.113...

Back-up and Error Memory

Perform a back-up and then erase the error memory in the generator and in the Multiline
(refer to Service Software Instructions, RXL2-120.034... and RXL2-120.113...) after
concluding the start-up.
Final Steps
If the shielding panels have been removed, reinstall them.
Attach the labels from register 9 (system binder) at appropriate locations.
Install all cover plates, except for the generator front panel and connect the protective
conductors.
Measure the protective conductor resistance.
Generator
Test meter:
Measurement
value:
Test procedure: The measurement is made between all conductive parts accessible to
Conditions:
Siemens AG
Medical Solutions
The input field "Actual:" is filled with blanks. These are not visible.
All banks must be deleted before entering the clinic name.
Prot. conductor meter
0.2 Ohm
Protective conductor meter
max. 0.2 Ohm (observe country-specific regulations)
(Current 10 A, voltage drop max. 2 V)
the user and the protective conductor rail in the generator.
Protective conductor resistance: max. 0.2 Ohm
If it is greater than this value, increase the wire gauge of the
protective conductor.
Reduce the transfer resistance (e.g. tighten the screws).