Service Software Installation And System Programming - Siemens LITHOSTAR Multiline SP Start-Up Instructions

Table of Contents

Advertisement

Start-Up of Multiline System
Fig. 1

Service Software Installation and System Programming

Install the POLYDOROS Service Software and System Programming on the Service PC
according to the Start-Up Instructions "Generator Configuration with the Service PC" for
POLYDOROS on LITHOSTAR Multiline (RXL2-120.034...).
- The System Programming is stored on the diskette FD-SW POLYDOROS in the
"Struct.dat" file.
- The diskettes are located in the Logbook.
Connect the service PC to the generator (Refer to chapter 1).
Switch on the system and the service PC.
NOTE
Prior to turning the system over to the customer, save the system programming on the
FD - SW POLYDOROS diskette (back-up).
Siemens AG
Medical Solutions
S1
The system was configured completely during the system test.
No changes may be made when the generator is started up. All
values measured must be entered in the generator test certificate.
RXL2-120.815.01
Rev. 02
Page 1 of 8
02.05
CS PS 24
4 - 1
4
D100
4
LITHOSTAR Multiline