Agilent E1364A Form C Switch Specifications
Maximum Input Voltage:
C to NC or NO:
250VDC
250VAC RMS
354VAC Peak
Maximum Current:
Per switch:
lA DC
lA AC RMS
Maximum Power:
Per switch:
40W DC
40VA AC
Per module:
320W DC
320VA AC
Thermal Offset: <7 V per chan.
Closed Channel Resistance:
< 1.5 /contact (initial)
< 3.5 at end of relay life
Insulation Resistance: (between any two points)
7
>10
at 40ºC, 65% RH
8
>10
at 25ºC, 40% RH
Bandwidth: -3dB at 10 MHz (typical)
Crosstalk, Channel to Channel: (typical)
100 kHz:
80dB
10 MHz:
30dB
Power Available:
+5V @ lA
+12V @ 0.5 A
Insertion Loss:
100kHz: 0.1dB
10MHz: 3.0dB
Appendix A
Any term. to chassis:
250VDC
250VAC RMS
354VAC Peak
Capacitance:
C to NC or NO: <20pF
Chan. to Chan.: <20pF
Relay Life (typical):
6
No load:
>10
operations
5
Max. load: >10
operations
Power Up/Down States: all open / left in last state
Terminals: screw type, max. wire size 16 AWG
Module Size/Device Type: B, register-based
Connectors Used: P1
No. Slots: 1
VXIbus Interface Capability: Interrupter, D16
Interrupt Level: 1-7, selectable
Power Requirements:
Voltage:
Peak module
current, IPM (A): 0.10 0.24
Dynamic module
current, IDM (A): 0.01 0.01
Watts/Slot: 1.0
Cooling/Slot: 0.02 mm H
Humidity: 65% to 40ºC
Operating Temperature: 0 to 55ºC
Storage Temperature: -40 to 75ºC
EMC, RFI, Safety: meets FTZ 1046/1984,
Net Weight (kg): 0.7
Agilent E1364A Form C Switch Specifications 63
Appendix A
+5
+12
0 @ 0.10 liter/sec
2
CSA 556B, IEC 348, UL 1244