90

Settings date and time

MAMMOMAT Stand

Prerequisite
The correct date and time must be set in the service PC. If not, set the date at C:\> by
typing date <ENTER> and the time at C:\ by typing time <ENTER>
Test performance
1. Start the Service Software for the MAMMOMAT Stand on the Service PC.
2. In the Main menu, select Configuration -> Clock.
Fig. 61: MAMMOMAT Stand clock adjustment
3. Press <F2>.
With <F2> the date and time set in the service PC are taken over by the MAMMOMAT
stand.

BRICK

1. Call up the WH AWS service software and select Utilities -> Escape to OS. Enter then
the following command in the Command field.
start telnet brick
2. A Username and Password have to be entered
Username: root
Password: ******
MAMMOMAT Novation DR
SPB7-250.815.06.02.02
12.05
Final procedures
Page 90 of 116
CS SD 24
0
0
0