Test 1 Protocol - Siemens MAMMOMAT Novation DR Manual

Hide thumbs Also See for MAMMOMAT Novation DR:
Table of Contents

Advertisement

64
2. Select the Procedure VendorProc 1 from the examination tab card.
Fig. 42: Select procedure for grid line test
NOTE
3. Select the exposure parameters: 28 kV and Mo/Mo.
4. Make sure that a 1 : 1 view is selected to have a better view.
Select View -> a 1:1
5. Make all exposures according to Tab. 1.
6. Activate magnification (Image -> Magnify by 2.0) and look for grid lines over the plexi
phantom area.
7. Make a not of the results (visible grid lines) in Tab. 1; choose the next image in the list
and start over with step 5 until all images are evaluated.

Test 1 protocol

Tab. 14
Checking the visibility of grid lines
Exposure
Exp Mode
1
2
3
MAMMOMAT Novation DR
Remember you can always add or remove images to the "Proce-
dure" by right-clicking the mouse.
mAs
mAs
32
mAs
56
mAs
110
SPB7-250.815.06.02.02
12.05
PMMA /
Tube Angle
mm
30
40
50
Page 64 of 116
CS SD 24
Image Quality
Visible grid lines
(yes/no)
o
0
o
0
o
0
Medical Solutions
0
Siemens AG