Final procedures

Settings date and time

MAMMOMAT Stand
Prerequisite
The correct date and time must be set in the service PC. If not, set the date at C:\> by
typing date <ENTER> and the time at C:\> by typing time <ENTER>.
Test performance
1.
2.
Fig. 1 MAMMOMAT Stand clock adjustment
3.

BRICK

Setting the date and time on the BRICK.
1.
2.
Siemens AG
Medical Solutions
Start the Service Software for the MAMMOMAT Stand on the Service PC.
In the Mainmenu select Configuration
Press <F2>.
With <F2> the date and time set in the service PC are taken over by the
MAMMOMAT stand.
Go into the service software and select Utilities -> Escape to OS and enter the
following command in the command field.
start telnet brick
A username and password have to be entered.
Username: root
Password: ******
Clock.