Service PC and measuring instruments

Saving the configuration file

NOTE
Main menu:
1. Select: Configuration -> Save config file.
2. Press F2 to save the data.
Reset the exposure counter
Main menu:
1. Select: Service -> Reset exposure counter.
2. Delete the error memory with Y.
Deleting the error buffer
Main menu:
1. Select: Service -> Delete error buffer.
2. Delete the error memory with Y.
Saving the MAMMOMAT Parameters
Main menu
1. Select Backup -> Copy Installation Area to Floppy -> Stand.
2. Select Backup -> Copy Installation Area to Floppy -> Panel.
3. Exit the program with F10 when the backup is completed.
NOTE
MAMMOMAT Novation DR
Make sure that the backup floppy is inserted.
Keep the backup floppy with other documentation for the MAM-
MOMAT Novation
SPB7-250.815.06.02.02
12.05
DR
in question.