Rosemount 3144P With Hart / 4-20 Ma; Approved Manufacturing Locations; European Union Directive Information; Hazardous Locations Installations - Emerson Rosemount 3144P Reference Manual

Temperature transmitter
Hide thumbs Also See for Rosemount 3144P:
Table of Contents

Advertisement

Reference Manual
00809-0100-4021, Rev EA
March 2007
Appendix B
ROSEMOUNT 3144P WITH HART / 4–20 mA
Approved Manufacturing
Locations
European Union
Directive Information
Hazardous Locations
Installations
www.rosemount.com
Product Certifications
Rosemount 3144P With HART / 4–20 mA . . . . . . . . . . . . . page B-1
Rosemount 3144P With F
Installation Drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page B-9
Rosemount Inc. – Chanhassen, Minnesota, USA
Rosemount Temperature GmbH – Germany
Emerson Process Management Asia Pacific – Singapore
The EC declaration of conformity for all applicable European directives for this
product can be found on the Rosemount website at www.rosemount.com. A
hard copy may be obtained by contacting your local sales representative.
ATEX Directive (94/9/EC)
Rosemount Inc. complies with the ATEX Directive.
Electro Magnetic Compatibility (EMC) (89/336/EEC)
EN 50081-1: 1992; EN 50082-2:1995; EN 61326-A1+A2+A3:1997 – Industrial
North American Certifications
Factory Mutual (FM) Approvals
I5
FM Intrinsic Safety and Non-incendive:
Intrinsically Safe for Class I/II/III, Division 1, Groups A, B, C, D, E, F, and
G.
Temperature codes: T4A (T
Zone Marking: Class I, Zone 0, AEx ia IIC
T4 (T
= – 50 to 60 °C)
amb
Intrinsically Safe when installed in accordance with control drawing
03144-0321.
Non-incendive for use in Class I, Division 2, Groups A, B, C, and D.
Suitable for use in Class II / III, Division 2, Groups F and G.
Non-incendive when installed in accordance with Rosemount drawings
03144-0321.
Temperature codes: T6 (T
Rosemount 3144P
fieldbus . . . . . . . . . page B-6
OUNDATION
= – 60 to 60 °C)
amb
T5 (T
= – 60 to 50°C)
amb
= – 60 to 60 °C),
amb
T5 (T
= – 60 to 85°C)
amb

Advertisement

Table of Contents
loading