Use With The Hart Tri-Loop - Emerson Rosemount 3144P Reference Manual

Temperature transmitter
Hide thumbs Also See for Rosemount 3144P:
Table of Contents

Advertisement

Rosemount 3144P
USE WITH THE HART
TRI-LOOP
3-24
NOTE
The 3144P is set to address 0 at the factory, allowing it to operate in the
standard point-to-point manner with a 4–20 mA output signal. To activate
multidrop communication, the transmitter address must be changed to a
number between 1 and 15. This change deactivates the 4–20 mA analog
output, sending it to 4 mA. The failure mode current also is disabled.
AMS
Right click and select "Configuration Properties" from the menu screen.
Select the "HART" tab. From here, assign the polling address.
To prepare the 3144P transmitter with dual-sensor option for use with a
Rosemount 333 HART Tri-Loop, you must configure the transmitter to Burst
Mode and set the process variable output order. In Burst Mode, the
transmitter provides digital information for the four process variables to the
HART Tri-Loop. The HART Tri-Loop divides the signal into separate 4–20 mA
loops for up to three of the following choices:
primary variable (PV)
secondary variable (SV)
tertiary variable (TV)
quaternary variable (QV)
When using the 3144P transmitter with dual-sensor option in conjunction with
the HART Tri-Loop, consider the configuration of the differential, average, first
good temperatures, Sensor Drift Alarm, and Hot Backup features (if
applicable).
NOTE
These procedures assume that the sensors and the transmitter are
connected, powered, and functioning properly, and that a Field Communicator
is connected to the transmitter control loop and is communicating
successfully. For communicator usage instructions, see "Commissioning" on
page 2-2.
Set the Transmitter to Burst Mode
To set the transmitter to burst mode, follow the steps below.
1. From the Home screen, select 1 Device setup, 3 Configuration, 4 Device
output configuration, 3 HART output, 4 Burst option to prepare to set the
transmitter to burst command 3. The communicator displays the Burst
option screen.
2. Select Process vars/current. The communicator returns to the HART
output screen.
3. Select 3 Burst mode to prepare to enable Burst Mode. The
communicator displays the Burst Mode screen.
4. Select On to enable Burst Mode. The communicator returns to the HART
output screen.
5. Select Send to download the new configuration information to the
transmitter.