580
END IF
590
OUTPUT @Mux;"OPEN (@"&Cc$&"00)"
600
OUTPUT @Mux;"CLOS (@"&Cc$&"15)"
610
OUTPUT @Dmm;"TRIG SGL"
620
ENTER @Dmm;Result (I,3)
630
IF Result (I,3) > .01 THEN
640
PRINT "Tree B path leakage out of tolerance ";Result$(I,3);" Volts"
650
END IF
660
OUTPUT @Mux;"*RST"
670
IF I < 5 THEN
680
PRINT "Test ";I+1;" complete"
690
PRINT "Turn power supply OFF"
700
PRINT "Press Continue for test ";I+2
710
PAUSE
720
END IF
730 NEXT I
740 PRINT "Leakage tests complete"
750 DISP "Press Continue to print measurement results"
760 PAUSE
770 CLEAR SCREEN
780 !
790 ! Print results
800 !
810 Format:IMAGE K,3X,D.DDDD," Vdc",3X,D.DDDD," Vdc",3X,D.DDDD,
" Vdc",3X,D.DDDD," Vdc"
820 PRINT
830 PRINT "
840 PRINT "
Tree B"
850 PRINT USING Format;"HI to LO
",Result(0,0),Result(0,1),Result(0,2),Result(0,3)
860 PRINT USING Format;"HI to Chassis
",Result(1,0),Result(1,1),Result(1,2),Result(1,3)
870 PRINT USING Format;"LO to Chasis
",Result(2,0),Result(2,1),Result(2,2),Result(2,3)
880 PRINT
890 PRINT "
900 PRINT "
Tree B"
910 PRINT USING Format;"HI to LO
",Result(3,0),Result(3,1),Result(3,2),Result(3,3)
920 PRINT USING Format;"HI to Chassis
",Result(4,0),Result(4,1),Result(4,2),Result(4,3)
930 PRINT USING Format;"LO to Chasis
",Result(5,0),Result(5,1),Result(5,2),Result(5,3)
940 END
Positive polarity leakage "
Direct
Channels
Negative polarity leakage "
Direct
Channels
Verification Tests 2-21
Tree A
Tree A

Advertisement