HI to Chassis Leakage
2-14 Verification Tests
Figure 2-7. Negative HI to LO Leakage Connections
1.
Make Hardware Connections
• Turn power supply and mainframe power OFF
• Connect DMM, power supply, and resistor as shown in
Figure 2-8
• Turn power supply and mainframe power ON
2.
Check Direct Terminals Leakage
• Send *RST to FET Multiplexer
• Send TRIG SGL to DMM
• Record the DMM reading in Table 2-1
Positive Polarity, HI to Chassis, Direct
• The DMM measurement should be less than 0.010 Vdc.
A measurement out of this range indicates a failure of the
FET Multiplexer