Agilent Technologies 75000 B Series Service Manual page 26

Fet multiplexer
Hide thumbs Also See for 75000 B Series:
Table of Contents

Advertisement

Example: Closed
Channel Resistance
2-8 Verification Tests
This example performs a closed channel resistance test of all measurement
paths. If a FET on resistance is >3.1 kΩ , the program prints a message
indicating which channel has failed. Before the closed channel
Test
measurement, the program checks for stuck channels. If a stuck channel is
found, the program prints a message and halts.
10!
RE-SAVE "CLOS_TEST"
20
ASSIGN @Dmm TO 722
30
ASSIGN @Mux TO 70914
40
DISP CHR$(129)
50
DIM Result(1,15),Tree(1,1),Path$(1)[4], Cc$[2],Ch$[2]
60
DATA HI,LO
70
READ Path$(*)
80
Cc$ = "01"
90
!
100
! Start test
110
!
120
CLEAR SCREEN
130
PRINT "Install Component Assembly and Test Fixture"
140
PRINT
150
PRINT " 1.
160
PRINT " 2.
170
PRINT " 3.
180
PRINT " 4.
190
PRINT " 5.
200
PRINT " 6.
210
PAUSE
220
!
230
! Measure closed channel resistance
240
!
250
FOR I = 0 TO 1
260
CLEAR SCREEN
270
PRINT TABXY(1,1), "Channel ";Path$(I);" to Direct ";Path$(I);"
Measurements"
280
PRINT TABXY(1,3),"Connect DMM Input HI lead to Channel
";Path$(I)
290
PRINT TABXY(1,4),"Connect DMM Input LO lead to Direct ";Path$(I)
300
DISP "Press Continue when connections are complete"
310
PAUSE
320
OUTPUT @Dmm;"PRESET NORM;FUNC OHM"
330
OUTPUT @Mux;"*RST"
340
!
350
! Check for stuck channels
360
!
! Card number
Turn mainframe and Agilent 3458A DMM power OFF"
Connect GPIB cable between mainframe and DMM"
Install component assembly into mainframe"
Attach test fixture to component assembly"
Turn mainframe and DMM power ON"
Press Continue when ready to begin testing"