The following table lists the bit definitions for the Questionable Data Register:
The overload bits are set once per
until a new INITiate is sent.
Bit
Bit Name
Number
0
Voltage Overload
1
Current Overload
2

Not Used

3
Not Used
4
Temperature Overload
5
Frequency
Overload/Underflow
6
Not Used
7
Not Used
8
Calibration Corrupt
9
Resistance Overload
10
Not Used
11
Lower Limit Failed
12
Upper Limit Failed
13
Not Used
14
Memory Overflow
15
Not Used
Agilent Truevolt Series DMM Operating and Service Guide
INITiate
command. If you clear an overload bit, it will not be set again
Decimal
Value
1
Only reported as event. Returns 0. Read the Event Register.
2
Only reported as event. Returns 0. Read the Event Register.
4
(Reserved for future use)
8
(Reserved for future use)
16
Only reported as event. Returns 0. Read the Event Register.
32
Only reported as event. Returns 0. Read the Event Register.
64
(Reserved for future use)
128
(Reserved for future use)
256
At least one calibration constant is corrupt.
512
Only reported as event. Returns 0. Read the Event Register.
1024
(Reserved for future use)
2048
The most recent measurement failed the lower limit test.
4096
The most recent measurement failed the upper limit test.
8192
(Reserved for future use)
16384
Reading memory is full. One or more (oldest) measurements
have been lost.
32768
(Reserved for future use)
Definition
225