Agilent Technologies Truevolt Series Operating And Service Manual page 211

Table of Contents

Advertisement

[SENSe:]TEMPerature Subsystem
This subsystem configures temperature measurements.
Command Summary
[SENSe:]TEMPerature:NPLC
[SENSe:]TEMPerature:NULL[:STATe]
[SENSe:]TEMPerature:NULL:VALue
[SENSe:]TEMPerature:NULL:VALue:AUTO
[SENSe:]TEMPerature:TRANsducer:{RTD|FRTD}:RESistance[:REFerence]
[SENSe:]TEMPerature:TRANsducer:{THERmistor|FTHermistor}:TYPE
[SENSe:]TEMPerature:TRANsducer:TYPE
[SENSe:]TEMPerature:ZERO:AUTO
[SENSe:]TEMPerature:NPLC {< PLC >|MIN|MAX|DEF}
[SENSe:]TEMPerature:NPLC? [{MIN|MAX|DEF}]
Sets the integration time in number of power line cycles (PLC) for temperature measurements. Integration time is the
period that the instrument's analog-to-digital (A/D) converter samples the input signal for a measurement. A longer
integration time gives better measurement resolution but slower measurement speed.
Parameter
{0.02|0.2|1|10|100}, default 10
See
Range, Resolution and NPLC
Configure 4-wire RTD measurements using a 10 PLC integration time. Then make and read one measurement:
CONF:TEMP FRTD
TEMP:NPLC 10
READ?
Typical Response: +6.27530000E+01
Only integration times of 1, 10, or 100 PLC provide normal mode (line frequency noise) rejection.
l
Setting the integration time also sets the measurement
l
between integration time and resolution.
This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet).
l
Agilent Truevolt Series DMM Operating and Service Guide
+1.00000000E+01
for more information.
Typical Return
resolution. Resolution Table
shows the relationship
211

Advertisement

Table of Contents
loading

This manual is also suitable for:

Truevolt 34460aTruevolt 34461a

Table of Contents