DC Current 10 A Full Scale Calibration (34461A only)
Review
Test Considerations
Configuration: DC Current
1. Configure the 10 A range and 100 PLC aperture.
2. Apply the 5 A input.
3. Enter the actual applied input current as the Cal Value (see
Succeeded message indicates success; if the display shows Calibration Step Failed, check the input value, range,
function, and entered adjustment value and repeat the adjustment step.
4. Store the new calibration constants (see
5.
Verify the AC Current Flatness
This calibration takes less than 4 seconds to complete.
344
and
Flatness Adjustment Considerations
Storing the Calibration
adjustments.
before beginning this procedure.
Entering Adjustment
Values). A Calibration Step
Constants).
Agilent Truevolt Series DMM Operating and Service Guide