Kenwood CS-4125 Instruciton Manual page 16

40/20mhz dual trace oscilloscope
Table of Contents

Advertisement

CHECKING AND ADJUSTMENT PRIOR TO MEASUREMENT
I n order to operate the oscilloscope at its optimum performance level, c a r r y out
the
following checks and adjustment before doing your measurements. T h e
instructions w h i c h follow concerning basic operation techniques and applications
assume that the checks and adjustment described here have been completed.
1.
Adjust the control panel to the following settings.
M O D E
S O U R C E
ffiaHJMMMODE
S L O P E
L E V E L
C H I ( Y ) ; C H 2 ( X )
^
P O S I T I O N
V A R I A B L E
V O L T S / D I V
A C - G N D - D C
H O R I Z O N T A L
< • P O S I T I O N
V A R I A B L E
S W E E P T I M E / D I V
X 1 0 M A G
N e x t , after c h e c k i n g the power source voltage r a t i n g s , s w i t c h the P O W E R
control on. T h e pilot lamp w i l l light up, and a trace line w i l l appear i n 10 to 15
seconds.
Check to see that rotating the I N T E N S I T Y control to the r i g h t
increases trace brightness, and rotating i t to the left decreases brightness.
T h e n rotate the I N T E N S I T Y control a l l w a y to left and e x t i n g u i s h the trace
line to being preheating. F o r the most accurate measurement results, i t i s
necessary to preheat the oscilloscope for about 30 m i n u t e s . H o w e v e r , i f you
intend only to display w a v e f o r m s , preheating is not necessary.
2.
After preheating, adjust the I N T E N S I T Y control so that the trace line is easy
to see, a n d adjust the F O C U S control to attain the clearest display image
possible. T h e n use the T R A C E R O T A control to b r i n g the trace line parallel
w i t h the horizontal graduation lines.
16
A U T O
V E R T M O D E
C H I ( I N V : O F F )
+
12 O ' C L O C K
12 O ' C L O C K
C A L
5 V / D I V
G N D
12 O ' C L O C K
C A L
0 . 2 m s / D I V
O F F