To configure a MSD transfer line heater

To configure a nickel catalyst heater

170
Scroll to Auxiliary type, press [Mode/Type], scroll to and
6
select the desired device type, and press [Enter]. Types
may include:
Cryo focus
Cryo trap
AED transfer line
Nickel catalyst
ICMPS argon preheat
ICMPS transfer line
ICPMS injector
Ion Trap GC Heated Interface
G3520 Transfer Line
MSD transfer line
User Configurable
7
When prompted, reboot the GC to implement the changes.
Check that a power source for the MSD heater was
1
assigned. See
"To assign a GC power source to an Aux
thermal zone"
on page 169.
Press [Config][Aux Temp #] and scroll to Thermal Aux 1,
2
Thermal Aux 2, or Thermal Aux 3 depending on where the
MSD heater was assigned, and press [Enter].
Scroll to Auxiliary type, press [Mode/Type], scroll to and
3
select the MSD transfer line, and press [Enter].
1
Check that a power source for the Nickel Catalyst heater
was assigned. See
Aux thermal zone"
Press [Config][Aux Temp #] and scroll to Thermal Aux 1,
2
Thermal Aux 2, or Thermal Aux 3 depending on where the
Nickel Catalyst heater was assigned, and press [Enter].
Scroll to Auxiliary type, press [Mode/Type], scroll to and
3
select Nickel catalyst, and press [Enter].
"To assign a GC power source to an
on page 169.