To Check the FID Baseline

Maintaining Your GC
1
With the column installed, load your checkout method.
2
Set the oven temperature to 35 °C.
Press [Front Detector] or [Back Detector], then scroll to
3
Output.
When the flame is lit and the GC is ready, verify that the
4
output is stable and < 20 pA.
If the output is not stable or > 20 pA, the system or gas
may be contaminated. If this contamination is isolated to
the detector, then bakeout the FID. (See
FID"
on page 229.)
11
Maintaining the FID
"To Bakeout the
225