Water capacity:
Standard water tub - material:
Cleanable water tub - material:
Temperature
Maximum heater plate:
Cut-out:
Maximum gas temperature:
Air filter
Standard:
Hypoallergenic:
Aircraft use
ResMed confirms that device meets the Federal Aviation Administration (FAA) requirements (RTCA/DO-160, section 21,
category M) for all phases of air travel.
Wireless module
Technology used:
FCC ID: 2ACHL-AIR10CD
The AirCurve 10 device complies with FCC Rules.
The AirCurve 10 device should be used at a minimum distance of 0.8" (2 cm) from the body during operation. Additional
information regarding the FCC Rules for this device can be found on www.ResMed.com/ProductSupport.
Operating pressure range
VAuto, S, ST, T:
CPAP:
Supplemental oxygen
Maximum flow:
Pneumatic flow path
Design life
Device, power supply unit:
Cleanable water tub:
Standard water tub, air tubing:
General
The patient is an intended operator.
To maximum fill line 380 mL
Injection molded plastic, stainless steel and silicone seal
Injection molded plastic, stainless steel and silicone seal
154°F (68°C)
165°F (74°C)
106°F ( 41°C)
Material: Polyester non woven fiber
Average arrestance: >75% for ~7 micron dust
Material: Acrylic and polypropylene fibers in a polypropylene
carrier
Efficiency: >98% for ~7-8 micron dust; >80% for ~0.5 micron
dust
CDMA (USA and Canada only)
2G GSM (all regions except USA and Canada)
3 to 25 cm H
O
2
4 to 20 cm H
O
2
15 L/min (S, ST, T, CPAP), 4 L/min (VAuto)
1. Flow sensor
2. Blower
3. Pressure sensor
4. Mask
5. Air tubing
6. Water tub
7. Device
8. Inlet filter
5 years
2.5 years
6 months