Safety guidelines

Related Links
Safety precautions

Recurrent safety tests

The printer shall be tested according to IEC 62353* in a time interval of at
least 36 months or less if local regulations are different.
*Medical electrical equipment – Recurrent test and test after repair of medical
electrical equipment.
Drystar 5301, Drystar 5302 | Preventive maintenance schedule | 107
WARNING:
To prevent damage to the printer while performing
maintenance, observe the following safety precautions:
• Do not lubricate the printer.
• Do not attempt to disassemble the printer.
• Do not touch the resistor line of the print head.
• Always switch off the printer and disconnect the power cord
from the outlet before carrying out any maintenance work
inside the printer.
Note: Film jam removal or cleaning the printer head can be done
without switching the power off.