Installation Manual
MMI-20027478
Variable
Frequency outputs (optional)
Discrete outputs (optional)
Discrete input (optional)
RS-485
Calibration parameters (for 9-wire installations only)
Flow calibration factor
Density calibration factors
2.3
Power requirements
Self-switching AC/DC input, automatically recognizes supply voltage:
•
85 to 240 VAC, 50/60 Hz, 6 watts typical, 11 watts maximum
•
18 to 100 VDC, 6 watts typical, 11 watts maximum
Note
For DC power:
•
Power requirements assume a single transmitter per cable.
•
At start-up, the power source must provide a minimum of 1.5 amps of short-term
current per transmitter and not pull voltage below 18 VDC.
•
Length and conductor diameter of the power cable must be sized to provide 18 VDC
minimum at the power terminals, at a load current of 0.7 amps.
Installation Manual
Setting
—
Power (nternal or external):
—
Source:
—
Scaling (LRV, URV):
—
Fault action:
—
Dual output:
—
Power (internal or external):
—
Source:
—
Scaling (LRV, URV):
—
Fault action:
—
Power (internal or external):
—
Source:
—
Scaling (LRV, URV):
—
Fault action:
Address:
FCF (Flow cal or flow calibration factor):
—
D1:
—
D2:
—
K1:
—
K2:
—
TC:
—
FD:
Planning
April 2022
9