2.2 Additional considerations for retrofitting existing
installations
The transmitter installation may require 3 in (76 mm) to 6 in (152 mm) of additional wiring for the input/
output and power connections. This length would be in addition to the currently installed wiring. Confirm
you have the additional wiring necessary for the new installation.
Before removing the existing transmitter, be sure to record the configuration data for the currently
installed transmitter. At initial startup of the newly installed transmitter, you will be prompted to
configure the meter via a guided setup.
Record the following information (if applicable):
Variable
Tag
Mass flow units
Volume flow units
Density units
Temperature units
Calibration parameters (for 9-wire installations only)
Flow calibration factor
Density calibration factors
Function block settings
Channel assignment
L_Type
XD_scale (engineering units
assignment)
Device address
Ident selection

2.3 Power requirements

Self-switching AC/DC input, automatically recognizes supply voltage:
85 to 240 VAC, 50/60 Hz, 6 watts typical, 11 watts maximum
18 to 100 VDC, 6 watts typical, 11 watts maximum
8
Setting
FCF (Flow Cal or Flow Calibration Factor):
D1:
D2:
K1:
K2:
TC:
FD:
Installation Manual
Micro Motion 5700 transmitters for PROFIBUS PA
MMI-20075112