General; Mammomat With Serial Number Lower Than 2091; Mammomat With Serial Number Within The Interval 2091-3241; Mammomat With Serial Number Within The Interval 3242-5199 - Siemens Opdima MAMMOMAT 1000 Installation And Start-Up Instructions Manual

Table of Contents

Advertisement

Prerequisites

General

Depending on the serial number of your MAMMOMAT, complementary measures may
have to be performed when installing Opdima?. Where necessary, the measures are
described in these instructions. The following has to be carried out with all serial numbers:
Installation of isolation board D707 (kit part No. 63 96 704) in the generator, if not already
done in conjunction with installation of printer (option)
The wall socket for the workstation power supply must be connected to the same branch
circuit as the MAMMOMAT power supply
This Installation and Start-Up Instruction is valid for Opdima? systems;
- with SUN workstation part No. 64 30 453 / Ultra 10 from serial number 1201

MAMMOMAT with serial number lower than 2091

The following has to be carried out:
Exchange of printed circuit board D702 (kit part No. 64 21 288) in the generator
Upgrade of the MAMMOMAT software (generator and stand) to V 2.2 (part No.
62 98 769, if the version to be upgraded is lower than V1.5, the AEC have to be
upgraded, see Update Instruction, UI 001/96)
Performance of "Stereotactic biopsy attachment update" according to MAMMOMAT
3000 Installation and Setting Instructions Update Kit for Stereotactic Biopsy Attachment
RX B7-230.033.02. ..

MAMMOMAT with serial number within the interval 2091-3241

The following has to be carried out:
Exchange of board D702 (kit part No. 64 21 288) in the generator
Upgrade of the MAMMOMAT software (generator and stand) to V2.2 (part No.
62 98 769, if the version to be upgraded is lower than V1.5, the AEC have to be
upgraded, see Update Instruction, UI 001/96)

MAMMOMAT with serial number within the interval 3242-5199

The following has to be carried out:
Upgrade of the MAMMOMAT software (generator and stand) to V2.2 (part No.
62 98 769, if the version to be upgraded is lower than V1.5, the AEC have to be
upgraded, see Update Instruction, UI 001/96)
Siemens-Elema AB
Solna, Sweden
Register 3
SPB7-230.033.10
Rev. 03
Page 1 of 4
08.02
SPS-UD
1 - 1
1
1
1
1
1
Opdima?
Installation