Restoring installation parameters

If the software has been changed, the installation parameters have to be restored.
Proceed as follows:
1.
2.
3.
4.
5.
NOTE
NOTE
6.
7.
8.
Opdima?
Installation
Switch on the MAMMOMAT. Install the service PC, (see Working with the Service
PC in MAMMOMAT 1000/3000/3000 Nova Installation and Start-Up Instruction).
Start the service program with the new service diskette.
Select Main menu => Backup => Copy disk to installation area => All. This will
copy the previously stored data from the hard disk to the installation area of the
MAMMOMAT.
Select Main menu => Configuration => Power and note the values for all four
focus.
Select Main menu => Configuration => Anode and enable/disable the W
(tungsten) anode depending on the system configuration.
For serial numbers lower than 5000 the W anode should always
be enabled.
For serial numbers higher than 5000 it depends on what the cus-
tomer ordered, if the W anode should be enabled or disabled. This
can be checked by looking at the filter disk in the collimator. If the
filter disk has only two filters and a lead strip mounted instead of
a third filter, the W anode should be disabled.
Select Main menu => Configuration => Power. Make sure that the values are
identical to the values noted in step 3. above, if not correct them.
Select Main menu => Configuration => Mischelaneous=> Cassette loaded switch
and set the switch to ON.
Switch off the MAMMOMAT with the main circuit breaker.
Register 3
SPB7-230.033.10
Rev. 03
08.02
Installation
Page 8 of 26
Siemens-Elema AB
SPS-UD
5
Solna, Sweden