Connecting the workstation

1.
CAUTION
2.
3.
4.
5.
Opdima?
Installation
Connect workstation cable marked X203 to the biopsy controller connector X203,
see Fig. 21, and the other end (X101) of the cable to the workstation connector
(10/Fig. 25).
Connecting the cable to the wrong connector may cause severe
damage to the equipment.
Attach ferrite on the cable on the biopsy controller side, see Fig. 21.
X201
X202
X203
Ferrite
X204
X205
X206
X207
Fig. 21 Biopsy controller
Connect the workstation cable marked X204 to the biopsy controller connector
X204 in Fig. 21 and the other end of the cable to the workstation connector
A (6A/Fig. 25) and B (6B/Fig. 25).
Attach ferrite on the cable on the biopsy controller side, see Fig. 21.
Mount the touch protection over the connectors X203 and X204.
-Place the plastic plate (1/Fig. 22) behind the connectors X203 and X204.
-Place the plastic cover (2/Fig. 22) on the other side of the connectors.
-Fasten the plastic screws (3/Fig. 22).
1
2
Fig. 22 Mounting the touch protection
Register 3
Protective earth cable
Mains cable
3
SPB7-230.033.10
Rev. 03
08.02
Installation
Biopsy controller
Page 20 of 26