Page 5
64 30 453. The workstation consists of the main components • - Hard Disk Drive - Camera Interface - Serial Interface • • Siemens-Elema AB Register 5 SPB7-230.061.02 Page 1 of 6 MAMMOMAT 3000 Opdima Solna, Sweden Rev. 02 05.99 SPS-UD Service...
Page 6
There are no more tools required than the below numbered unless otherwise noted. • anti-static wristband and grounded tabletop pad (TDK equipment Type 8005 from 3M Deutschland GmbH). • screwdriver • Phillips screwdriver MAMMOMAT 3000 Opdima Register 5 SPB7-230.061.02 Page 2 of 6 Siemens-Elema AB Service Rev. 02 05.99...
Page 7
Select "shutdown" in the "login" dialogue. NOTICE The power is turned off automatically when shutdown is selected. Top Cover Fig. 2 Removing the top cover Siemens-Elema AB Register 5 SPB7-230.061.02 Page 3 of 6 MAMMOMAT 3000 Opdima Solna, Sweden Rev. 02 05.99 SPS-UD Service...
Page 8
Monitor Interface Serial Line B (Biopsy Contr. X204) Keyboard Interface SCSI Interface <...> SCSI Interface Tab. 1 Rear Side of the Ultra 1 with all Interfaces MAMMOMAT 3000 Opdima Register 5 SPB7-230.061.02 Page 4 of 6 Siemens-Elema AB Service Rev. 02 05.99...
Page 9
Hold the sides of the cover and lift it from the back panel. When the cover clears the back panel, move the cover away from the front panel to free it from the chassis. Siemens-Elema AB Register 5 SPB7-230.061.02 Page 5 of 6 MAMMOMAT 3000 Opdima Solna, Sweden Rev. 02 05.99 SPS-UD Service...
Page 10
This page intentionally left blank. MAMMOMAT 3000 Opdima Register 5 SPB7-230.061.02 Page 6 of 6 Siemens-Elema AB Service Rev. 02 05.99 SPS-UD Solna, Sweden...
Page 11
Loosen all four screws form the back of the box (refer to Fig. 2). Screws C and D secure the CPU to the back plate of the box. Siemens-Elema AB Register 5 SPB7-230.061.02 Page 1 of 20 MAMMOMAT 3000 Opdima Solna, Sweden Rev. 02 05.99 SPS-UD Service...
Page 12
Power Supply SBUS Memory, SIMM SCSI Disk 0 SPARC CPU Ultra 1 Fig. 3 Connectors on the Ultra 1 OPD00276 FRONT MAMMOMAT 3000 Opdima Register 5 SPB7-230.061.02 Page 2 of 20 Siemens-Elema AB Service Rev. 02 05.99 SPS-UD Solna, Sweden...
Page 13
Remove the Frame Buffer (refer to the chapter "Replacing the Frame Buffer"). • On the new CPU board, install the previously removed NVRAM and the memory modules. Siemens-Elema AB Register 5 SPB7-230.061.02 Page 3 of 20 MAMMOMAT 3000 Opdima Solna, Sweden Rev. 02 05.99 SPS-UD Service...
Page 14
• Mount the top cover of the box (refer to chapter 1). • Switch on the system. • Ensure that the system is operating properly. MAMMOMAT 3000 Opdima Register 5 SPB7-230.061.02 Page 4 of 20 Siemens-Elema AB Service Rev. 02 05.99...
Page 15
Remove the frame buffer board and place it on the grounded tabletop pad. • Locate the NVRAM and remove the NVRAM from the board (refer to Fig. 5). Siemens-Elema AB Register 5 SPB7-230.061.02 Page 5 of 20 MAMMOMAT 3000 Opdima Solna, Sweden Rev. 02 05.99 SPS-UD Service...
Page 16
Change the parameter values of the NVRAM to the correct values. • Replace the old Internet address with the new one. • Replace the old CPU no. with the new one. MAMMOMAT 3000 Opdima Register 5 SPB7-230.061.02 Page 6 of 20 Siemens-Elema AB Service Rev.
Page 17
Siemens-Elema AB Register 5 SPB7-230.061.02 Page 7 of 20 MAMMOMAT 3000 Opdima Solna, Sweden Rev. 02 05.99 SPS-UD Service...
Page 18
Read these instructions completely prior to beginning the replacement. Follow all procedures for removing and installing the DSIMM. Follow all general safety guidelines when replacing the DSIMM. Note the special safety guidelines indicated in these instructions. MAMMOMAT 3000 Opdima Register 5 SPB7-230.061.02 Page 8 of 20...
Page 19
Fig. 7). • Release the module from the socket and place it on a grounded tabletop pad. Fig. 7 SIMM retainers Siemens-Elema AB Register 5 SPB7-230.061.02 Page 9 of 20 MAMMOMAT 3000 Opdima Solna, Sweden Rev. 02 05.99 SPS-UD Service...
Page 20
Switch on power to the system. • Ensure that the system is functioning properly. • After the reset the system should show up with: "64 MB Memory installed". MAMMOMAT 3000 Opdima Register 5 SPB7-230.061.02 Page 10 of 20 Siemens-Elema AB Service Rev.
Page 21
Sun Ultra 1 UPA / SBus (Ultra SPARC 143 MHz), Keyboard Present Open Boot 3.3, 64 MB Memory installed, Serial#..Ethernet address 8:.., Host ID:..Fig. 9 System Message Siemens-Elema AB Register 5 SPB7-230.061.02 Page 11 of 20 MAMMOMAT 3000 Opdima Solna, Sweden Rev. 02 05.99 SPS-UD Service...
Page 22
FRONT Fig. 10 Location of Camera Interface D100 Read these instructions completely prior to beginning the replacement. Follow all procedures for removing and installing the Camera Interface D100. MAMMOMAT 3000 Opdima Register 5 SPB7-230.061.02 Page 12 of 20 Siemens-Elema AB Service Rev.
Page 23
• Remove the board and place it on the grounded tabletop pad. Fig. 11 Card retainers Siemens-Elema AB Register 5 SPB7-230.061.02 Page 13 of 20 MAMMOMAT 3000 Opdima Solna, Sweden Rev. 02 05.99 SPS-UD Service...
Switch on power to the system. • You may check the 5 V power on the test points (refer to Fig. 16). • Ensure that the system is operating properly. MAMMOMAT 3000 Opdima Register 5 SPB7-230.061.02 Page 14 of 20 Siemens-Elema AB Service Rev.
Page 25
CIA card is a credit-card-sized, removable module that has become the expansion vehicle for (portable) computers. In the OPDIMA system the PCMCIA Interface is used for the connection to a MODEM. Siemens-Elema AB Register 5 SPB7-230.061.02 Page 15 of 20 MAMMOMAT 3000 Opdima Solna, Sweden Rev. 02 05.99 SPS-UD Service...
Page 26
Carefully lift the corners of the board to release it from the connector • Remove the board and place it on the grounded tabletop pad. Fig. 14 Card retainers MAMMOMAT 3000 Opdima Register 5 SPB7-230.061.02 Page 16 of 20 Siemens-Elema AB Service Rev. 02 05.99...
Eject button Fig. 15 PCMCIA Interface • Switch on power to the system. • Ensure that the system is operating properly. Siemens-Elema AB Register 5 SPB7-230.061.02 Page 17 of 20 MAMMOMAT 3000 Opdima Solna, Sweden Rev. 02 05.99 SPS-UD Service...
Page 28
Follow all procedures for removing and installing the Frame Buffer board. Follow all general safety guidelines when replacing the Frame Buffer. Note the special safety guidelines indicated in these instructions. MAMMOMAT 3000 Opdima Register 5 SPB7-230.061.02 Page 18 of 20...
Page 29
Insert the new Frame Buffer carefully in the original location. • Check the jumpers of the frame buffer (refer to Fig. 17). Fig. 17 Jumper Setting Frame Buffer Siemens-Elema AB Register 5 SPB7-230.061.02 Page 19 of 20 MAMMOMAT 3000 Opdima Solna, Sweden Rev. 02 05.99 SPS-UD Service...
Page 30
• Switch on power to the system. • Ensure that the system is operating properly. MAMMOMAT 3000 Opdima Register 5 SPB7-230.061.02 Page 20 of 20 Siemens-Elema AB Service Rev. 02 05.99 SPS-UD Solna, Sweden...
Page 31
SCSI ID. Distribution Disk SCSI ID CD-drive CD-drive installed when SW-upgrade Data Disk SCSI ID MO-drive Fig. 1 Disk System Siemens-Elema AB Register 5 SPB7-230.061.02 Page 1 of 10 MAMMOMAT 3000 Opdima Solna, Sweden Rev. 02 05.99 SPS-UD Service...
Page 32
To detect a faulty device enter ok probe-scsi Fig. 2 Test of the SCSI bus The system should show all the drives connected to the bus: MAMMOMAT 3000 Opdima Register 5 SPB7-230.061.02 Page 2 of 10 Siemens-Elema AB Service Rev.
Page 33
Swing the plastic latch to the left and slide the disk drive support and disk drive outward on the guide rails and remove it from the box. Siemens-Elema AB Register 5 SPB7-230.061.02 Page 3 of 10 MAMMOMAT 3000 Opdima Solna, Sweden Rev. 02 05.99 SPS-UD Service...
Page 34
• Swing the plastic latch to the left and Side View OPENER MAMMOMAT 3000 Opdima Register 5 SPB7-230.061.02 Page 4 of 10 Siemens-Elema AB Service Rev. 02 05.99 SPS-UD Solna, Sweden...
Page 35
* CD ROM Disk Drive is installed when SW upgrade shall be performed. System Disk SCSI ID Seagate ST32550WC J2 connector Pin 1 J4 connector SCSI ID = 0 Pin 1 Siemens-Elema AB Register 5 SPB7-230.061.02 Page 5 of 10 MAMMOMAT 3000 Opdima Solna, Sweden Rev. 02 05.99 SPS-UD Service...
Page 36
The system should show all the drives connected to the bus (refer to Fig. 3). • If the disk drives shows up correctly load the software (refer to chapter "Loading the System Software"). See MAMMOMAT 3000 Opdima™ Digital Biopsy and Spot Imaging System - Service Instructions (SPB7-230.061.03..) MAMMOMAT 3000 Opdima Register 5 SPB7-230.061.02...
(refer to Fig. 10). • Unplug the SCSI terminator from the back of the unit (refer to Fig. 10). • Remove the unit. Siemens-Elema AB Register 5 SPB7-230.061.02 Page 7 of 10 MAMMOMAT 3000 Opdima Solna, Sweden Rev. 02 05.99 SPS-UD Service...
Page 38
* CD ROM Disk Drive is installed when SW upgrade shall be performed. In/Out SCSI SCSI ID Switch Device Mode Switch connectors AC in ON/OFF switch Fig. 9 MO System, back view MAMMOMAT 3000 Opdima Register 5 SPB7-230.061.02 Page 8 of 10 Siemens-Elema AB Service Rev. 02 05.99 SPS-UD...
SCSI terminator Fig. 10 Archive System SCSI ID Switches • Switch on the system. • Ensure that the system is operating properly. Siemens-Elema AB Register 5 SPB7-230.061.02 Page 9 of 10 MAMMOMAT 3000 Opdima Solna, Sweden Rev. 02 05.99 SPS-UD Service...
Page 40
This page intentionally left blank. MAMMOMAT 3000 Opdima Register 5 SPB7-230.061.02 Page 10 of 10 Siemens-Elema AB Service Rev. 02 05.99 SPS-UD Solna, Sweden...
Page 41
Disconnect the power supply cable from the back of the box. • Loosen the screw that secures the power supply (refer to 1, Fig. 1). Siemens-Elema AB Register 5 SPB7-230.061.02 Page 1 of 4 MAMMOMAT 3000 Opdima Solna, Sweden Rev. 02 05.99 SPS-UD Service...
Page 42
Plug 2 Plug 3 Fig. 3 Power Supply Connector • Move the power supply a few millimeters to the left side to release it from the bottom. MAMMOMAT 3000 Opdima Register 5 SPB7-230.061.02 Page 2 of 4 Siemens-Elema AB Service Rev.
Page 43
Connect the power supply cable to the back of the box. • Switch on the system. • Ensure that the system is operating properly. Siemens-Elema AB Register 5 SPB7-230.061.02 Page 3 of 4 MAMMOMAT 3000 Opdima Solna, Sweden Rev. 02 05.99 SPS-UD Service...
Page 44
This page intentionally left blank. MAMMOMAT 3000 Opdima Register 5 SPB7-230.061.02 Page 4 of 4 Siemens-Elema AB Service Rev. 02 05.99 SPS-UD Solna, Sweden...
Page 45
Note the special safety guidelines indicated in these instructions. The entire system should be switched off when the monitor is being replaced. NOTICE Fig. 2 Symbol Siemens-Elema AB Register 5 SPB7-230.061.02 Page 1 of 6 MAMMOMAT 3000 Opdima Solna, Sweden Rev. 02 05.99 SPS-UD Service...
Page 46
Rear Side of the box, Monitor Interface • Remove the monitor. Install the monitor by reversing the order in which it was removed. • Ensure that the system is operating properly. MAMMOMAT 3000 Opdima Register 5 SPB7-230.061.02 Page 2 of 6 Siemens-Elema AB Service Rev.
Page 47
Keyboard Interface Fig. 4 Back of the box (without D100). CAUTION • Remove the keyboard connector in back of the box. Siemens-Elema AB Register 5 SPB7-230.061.02 Page 3 of 6 MAMMOMAT 3000 Opdima Solna, Sweden Rev. 02 05.99 SPS-UD Service...
Page 48
NOTICE • Remove keyboard with cable. Install the new keyboard by reversing the order in which it was removed. • Ensure that the system is operating properly. MAMMOMAT 3000 Opdima Register 5 SPB7-230.061.02 Page 4 of 6 Siemens-Elema AB Service Rev.
Page 49
The entire system should be switched off when parts are being replaced. CAUTION • Remove the mouse with the cable (refer to Fig. 6). Siemens-Elema AB Register 5 SPB7-230.061.02 Page 5 of 6 MAMMOMAT 3000 Opdima Solna, Sweden Rev. 02 05.99 SPS-UD Service...
Page 50
Install the new mouse by reversing the order in which it was removed. Fig. 6 Rear view of the keyboard: Installing the mouse • Ensure that the system is operating properly. MAMMOMAT 3000 Opdima Register 5 SPB7-230.061.02 Page 6 of 6...
Page 51
System Disk........3 - 3 System Software......3 - 6 top cover .......... 2 - 1 Ultra 1 ........2 - 1, 2 - 3 Siemens-Elema AB Register 5 SPB7-230.061.02 Page 1 of 2 MAMMOMAT 3000 Opdima Solna, Sweden Rev. 02 05.99 SPS-UD Service...
Page 52
This page intentionally left blank. MAMMOMAT 3000 Opdima Register 5 SPB7-230.061.02 Page 2 of 2 Siemens-Elema AB Service Rev. 02 05.99 SPS-UD Solna, Sweden...