Maintenance; Cleaning; Tubing Inspection - Teledyne CombiFlash Companion XL Installation Manual

Table of Contents

Advertisement

CombiFlash

9.1 Cleaning

For general cleaning of the enclosure, use distilled water with
a mild detergent. Use isopropyl alcohol for tougher stains.
On printed areas, such as the rear panel, avoid rubbing
vigorously or using aggressive solvents like acetone. Each will
ruin the printed text.
Do not immerse the instrument in a water bath. The instrument is
not watertight and this action could damage the internal
electronics.

9.2 Tubing Inspection

Risk of fire or equipment damage. Faulty tubing, fittings, and
drains may allow organic solvents to pool in unsafe areas,
creating a potential for dangerous levels of flammable vapors.
Visually inspect the solvent, waste, and drain tubing monthly.
The tubing must be free of any damage, kinks, or
deterioration. Fittings should show no signs of leaks.
The top shelf drain and Foxy 200 drain should also be
inspected at this time. These drain holes should be clear of any
debris.
Correct any deficiencies before returning the instrument to
operation.
®
Companion
Installation Guide
Section 9 Maintenance
CAUTION
WARNING
®
XL
9-1

Advertisement

Table of Contents
loading

Table of Contents