Test Limits for PASS result
Earth Bond 200mA
Earth Bond 25A
Earth Bond 10A
Insulation Class I
Insulation Class II
Substitute Leakage Class I
Substitute Leakage Class II
Leakage Current
Touch Current
IEC lead Earth Bond
IEC lead Insulation
Variation Factor Errors
Variation Factor
Position
Supply Voltage
Temperature
Current Consumption
Magnetic Fields
Impedance
Capacitance
Current Waveshape
UK
<0.10 Ω
<0.10 Ω
NA
>1 MΩ
>2 MΩ
<3.5 mA
<0.50 mA
<0.75 mA
>0.25 mA
<0.10 Ω
>2 MΩ
Designation
E1
0.0 %
E2
5.0 %
E3
5.5 %
E4
1.5 %
E5
2.5 %
E6
1.0 %
E7
2.0 %
E8
1.0 %
AU
DE
<1.0 Ω
<0.30 Ω
<1.0 Ω
NA
NA
<0.30 Ω
>1 MΩ
>1 MΩ
>1 MΩ
>2 MΩ
<1.0 mA
<1.0 mA
<1.0 mA
<0.50 mA
<5.0 mA
<3.5 mA
>1.0 mA
<0.50 mA
<1.0 Ω
<0.30 Ω
>1 MΩ
>1 MΩ
% Variation Error
Appliance Tester
Specifications
NL
<0.30 Ω
<0.30 Ω
NA
>1 MΩ
>2 MΩ
<1.0 mA
<0.50 mA
<3.5 mA
<0.50 mA
<0.30 Ω
>1 MΩ
25