2-Ohm Output Current Test; Open Circuit Voltage Test - Fluke 150 Series Calibration Manual

Insulation testers
Hide thumbs Also See for 150 Series:
Table of Contents

Advertisement

5. Press T, and verify that the UUT reading is within the display reading limits
shown in Table 9.
6. Connect the Fluke 5320A high resistance output terminals to the UUT
Ω
terminals.
7. Set the Fluke 5320A to the high resistance source function (H).
8. Press T, and verify that the reading is within the display reading limits for
Table 9, Step 4.
Step
Tester Range
20.00 Ω
1
2000 Ω
2
2000 Ω
3
4
20.00 kΩ

2-Ohm Output Current Test

The following procedure verifies the minimum current for a 2-Ω Load in continuity
function:
1. Connect the Fluke 5320A low resistance output terminal to the UUT
terminals.
2. Set the Fluke 5320A to the low resistance source function (L) and 2-Ω output.
3. Set the UUT rotary function switch to the Ω function.
4. Press the UUT T. The Fluke 5320A current reading should be > 200 mA.

Open Circuit Voltage Test

The following test confirms that the open circuit voltage for the UUT Ω function is
within limits.
1. Connect the Fluke 5320A
terminals observing correct polarity.
COM
2. Set the Fluke 5320A to the M mode and dc volts.
3. Set the UUT rotary function switch to the
4. Press T and verify that the Fluke 5320A voltage reading is > 4.0 V but
< 8.0 V.
riešenia na presné meranie
Table 9. Earth Bond Resistance Tests
Applied
Display Units
2.0 Ω
810.0 Ω
990.0 Ω
18.0 kΩ
and
V
COM
http://www.elso.sk
Display Reading
Low Limit
Ω
1.94
Ω
795
Ω
972
17.70
meter terminals directly to the UUT
Ω
function.
Insulation Testers
Performance Tests
and
COM
High Limit
2.06
825
1008
18.30
and
Ω
COM
Ω
and
21