Load Current Test - Fluke T+ Calibration Information Manual

Electrical tester
Hide thumbs Also See for T+:
Table of Contents

Advertisement

Step
Function
14
24 V
15
120 V
16
208 V
17
240 V
Voltage LEDs
18
277 V
347 V
19
(Canada)
20
480 V
21
Continuity
0 – 20 kΩ
22
23
Resistance
0.00 – 9.99 kΩ
24

Load Current Test

To test the load current, set up the UUT with equipment as shown in Figure 2 and complete the
following procedure:
1. Set the DMM to the mA dc function.
2. Apply 600 V dc from the Source. The DMM reading should be < 5 mA.
3. Apply −600 V dc from the Source. The DMM reading should be < 5 mA.
Table 3. T+ and T+ PRO Accuracy Tests (cont.)
Input
Range
Value
24 V
120 V
208 V
240 V
277 V
347 V
480 V
Short
Short
100 Ω
9 kΩ
UUT Response
(LED)
24 V LED On
120 V LED On
208 V LED On
240 V LED On
277 V LED On
347 V LED On
480 V LED On
Continuity LED On
None
None
None
Electrical Tester
Performance Tests
LCD Display (T+ PRO)
Measurement Limits
- - -
- - -
- - -
- - -
- - -
- - -
- - -
Continuity Sign
0.00 to 0.03
0.06 to 0.14
8.52 to 9.48
9