Fluke 6200-2 User Manual page 32

Appliance tester
Table of Contents

Advertisement

6200-2
Users Manual
Load/ Leakage Test: Load Current
Display Range
UK ................................................................................... 0 A to 13 A
AU ................................................................................... 0 A to 10 A
DE, NL ............................................................................ 0 A to 16 A
Accuracy ................................................................................. ±(4 % + 2 counts)
Resolution ............................................................................... 0.1 A
The appliance under test is energized at mains potential.
Load/Leakage Test: Load Power
Display Range 230 V mains
UK ................................................................................... 0.0 VA to 3.2 kVA
AU ................................................................................... 0.0 VA to 2.4 kVA
DE, NL ............................................................................ 0.0 VA to 3.7 kVA
Accuracy ................................................................................. ±(5 % + 3 counts)
Resolution ............................................................................... 1 VA (0 to 999 VA), 0.1 kVA (>1.0 kVA)
The appliance under test is energized at mains potential.
Load/Leakage Test: Leakage Current (I
Operational Error Measurement Range .................................. 0.25 to 19.00 mA
Operational error ..................................................................... 12.0 %
Accuracy ................................................................................. ±(4 % + 5 counts)
Display Range ......................................................................... 0.25 to 19.99 mA
Resolution ............................................................................... 0.01 mA
The appliance under test is energized at mains potential.
IEC Lead Test
Test current ............................................................................. 10 A ac
Test voltage ............................................................................ 500 V dc
PELV Test
Display .................................................................................... "> PEL" indicator only
Accuracy at 50 Hz ................................................................... ±(2 % + 3 counts)
Overload protection ................................................................. 300 Vrms
Warning threshold ................................................................... 25 Vrms