Chapter 2
Operating the NI 435x Device

Knowing the Signal Source

Floating Signal Source

Ground-Referenced Signal Source

Using Programmable Ground-Referencing

NI 435x User Manual
For accurate measurements, you must determine whether the signal source
is floating or ground-referenced.
A
floating signal source
the building ground system but has an isolated ground-reference
point. Examples of floating signal sources are thermocouples with
ungrounded junctions and outputs of transformers, batteries,
battery-powered devices, optical isolators, and isolation amplifiers.
A
ground-referenced signal source
the building system ground. Therefore, it is already connected to a common
ground point with respect to the NI 435x, assuming that the computer is
plugged into the same power system. Examples of ground-referenced
signal sources include the following:
Thermocouples with grounded or exposed junctions connected to
grounded test points
The outputs of plug-in devices with nonisolated outputs
Voltage across RTDs, thermistors, or resistors that you may be
measuring using the built-in current sources of the NI 435x
The NI 435x devices have software-programmable ground-referencing on
every channel, which you can use to ground-reference a floating signal
source. This connects CH– to ground through a 10 MΩ resistor and
provides a ground-reference for the floating signal source. Even if the
signal source is ground-referenced, this resistance minimizes the effects of
ground-loops, as long as the source impedance and the lead wire resistance
is less than 100 Ω . Thus, you can take accurate measurements even if you
are uncertain whether the signal source is floating or ground-referenced.
Because you can set ground-referencing on a channel-by-channel basis,
you can have ground-referenced signal sources connected to some channels
and floating signal sources connected to other channels in the same
measurement setup. Table 2-2 summarizes the settings to use for
ground-referencing.