System Specifications And Characteristics - Agilent Technologies 34980A Service Manual

Multifunction switch/measure unit
Hide thumbs Also See for 34980A:
Table of Contents

Advertisement

2
Specifications

System Specifications and Characteristics

Scanning inputs
Analog
Digital
Scan triggering
Source
Scan count
Scan interval
Channel delay
External trig delay
External trig jitter
Alarms
Analog inputs
Digital inputs
Monitor channel
Alarm outputs
Latency
Memory
Type
Readings
States
Alarm queue
System features
Per-channel math 
min/max/average
Power fail recovery
Relay maintenance
Real-time clock
36
34921A, 34922A, 34923A, 34924A, and 34925A multiplexer
channels
34950A/52A digital in and totalize
Interval, external, button press, software, or on monitor channel
alarm
1 to 50,000 or continuous
0 to 99 hours; 1 ms step size
0 to 60 seconds per channel; 1 ms step size
< 2 ms. With monitor on < 200 ms
< 2 ms
Hi, Lo, or Hi + Lo evaluated each scan
34950A/52A digital in maskable pattern match or state change
34950A/52A frequency and totalize: Hi limit only
Alarm evaluated each reading
4 TTL compatible
Selectable TTL logic Hi or Lo on fail
5 ms (typical)
Volatile
500,000 with timestamp, readable during scan
5 instrument states with user label
Up to 20 events with channel number, reading, and timestamp
Individual Mx+B scaling and calculated real time
Save switch states
Counts each relay closure and stores on module, 
user resettable
Battery-backed, 20-year typical life
34980A Service Guide