SPECIFICATIONS
Table 3. Specifications
FUNCTION
RANGE
RESOLUTION
ACCURACY'
3.200V
32.00V
320.0V
1000V
0.001V
0.01V
0.1V
1V
+/-(0.1%+1)
+/
- (0.1%+1)
+/
- (0.1%+1)
+/-(0.
1
%+1)
mV
320.0 mV
0.1 mV
+/
- (0.1%+1)
320.00
0.10
3.200 kcl
0.001 Ic0
+/-(0.25%+1)
32.00 1(0
0.01 kla
+1-(0.25%+1)
320
. o
0.1 Ida
+/-(0.25%+1)
3.200 MO
0.001 MO
+/-(0.25%+1)
32.00 MO
0.01 MO
+/-(1%+1)
(nS)
32.00 nS
0.01 nS
+/-(2%+10) typical
Vw
41-
2.000V
0.001V
+/-(1%+1) typical
For continuity tests, the beeper emits a continuous tone at
test resistance below approximately 1500
20 Hz-40 Hz
40 Hz-1 kHz
1 kHz-5 kHz
3.200V
0.001V
+1-(1.5%+5)
+/-(5%+5)
32.00V
0.01V
+/-(1.5%+5)
+/-(0.5%+5)
320.0V
0.1V
+/-(1.5%+5)
+1-(0.51+5)
+/-(51+5)
1000V
1V
+1-(1.5%+5)
rat
320.0 mV
0.1 mV
+/-(1.5%+5)
+/-(0.5%+5)
+/-(5%+5)
* Accuracy is specified as +/-C% of reading] + [number of least
significant digits]).
Basic electrical accuracy is specified from 18 to 28 °C, at a relative
humidity of up to
95%,
for one year after calibration.
Within a single ac range, accuracy is specified between
5
and 100% of
full range. AC conversion method is AC coupled, true rms responding to
signals whose crest factor is 3:1 or less.