...8 SENSOR MAINTENANCE
...8.2.1 Dismantling and Cleaning – Figs. 8.1 to 8.4
Unscrew the collar and
carefully withdraw the
receiver module.
3
Split the flowcell – see Fig. 8.3.
Clean the inside of the flowcell
4
and other assemblies using mild detergent.
Rinse thoroughly with de-ionized water.
Inspect the wiper blades for wear or damage.
5
Replace as necessary – see Fig. 8.4.
6
Reassemble the flowcell using new seals – see Fig. 8.3.
Refit the emitter and receiver modules using new seals,
7
ensuring that the module locating pins are seated correctly
in their keyways before tightening the collars.
52
Sample Outlet
1
Fig. 8.2 7320 200 High Range Sensor – Dismantling and Cleaning
Unscrew the collar and carefully
2
withdraw the emitter module.
Inlet/Isolating
Valve
Flowcell
Drain
Valve
Seal
Seal