5.5 Configure Outputs

-----
CONFIG. OUTPUTS
-----
Config. Output 1
-----
Sen. A
AO1: Assign
0-20mA
0-10mA
-----
4-20mA
AO1: Range
28. 0 0
°H
14. 0 0
mg/l C
AO1: Span Value
Hold
On
-----
Off
mA
AO1: Default
Config. Output 1
Off
or
Hold
On
12. 0 0
mA
AO1: Default Val
Config. Output 1
Configure Output 1
Output 2 configuration is identical to Output 1 configuration.
Config. Output 2
Assign
Select either Sensor A (
Sen. A
Range
Select the analog output current range for the selected output.
Span Value
Inferred units. Displayed only if
Adjust the displayed reading to the required span value. Low range between 5.00 and
20.00 and high range between 20.0 and 100.0
Default Output
Select the system reaction to failure:
Off
– Ignore failure and continue operation.
On
– Stop on failure. This drives the analog output to the level set in the
frame below.
Hold
– Hold the analog output at the value prior to the failure.
Default Value
The level to which the analog output is driven if a failure occurs.
Set the value, between 0.00 and 22.00mA.
Configure Output 2.
Config. Output 2
See Section 5.6.
CONFIG. CLOCK
) or Sensor B (
).
Sen. B
set to anything other than
Infer. Units
5 PROGRAMMING...
– see Section 5.3.
None
Default Val