...4 SETUP
...4.2 Sensor Calibration
50. 0
mg/l C
50. 0
mg/l C
A:Fill Span Sol.
25.0
50. 0
R100.0
100%
50. 0
mg/l C
A:Cal. Complete
Calib. Sensor A
16
Span Calibration
Fill the flowcell with the required span standard calibration solution – see Section 4.1.2.
Press the
key to initiate calibration.
Note. If the
key is pressed again at any time before span calibration is complete,
is shown on the lower display line (see below).
A:Cal Incomplete
Raw Output from Sensor
Organics total signal
Reference total signal
As calibration proceeds, a progress indicator appears in the lower display line. After 3
minutes, the lower display line shows
automatically to the next frame.
Calibration Completed
A message is displayed at the end of calibration:
– calibration successful
A:Cal Complete
– zero and/or span calibration aborted
A:Cal Incomplete
Sensor B calibration (dual input monitors only) is identical to Sensor A
Calib. Sensor B
calibration.
Single input monitors only – return to main menu.
SENSOR CAL
Alter Sec. Code
SECURITY CODE
CONFIG. DISPLAY
Alter Sec. Code
These values equate to the percentage light transmission
through the sample.
100%
not set to zero (Section 5.8) – see Section 5.1.
set to zero (Section 5.8) – see Section 5.2.
, the display then advances