ABB AV410 User Manual page 50

Single and dual input dissolved organics monitor
Table of Contents

Advertisement

...7 CALIBRATION
...7.1 Factory Settings
-----
Cal. Output 1
16000
O1: Adjust 4mA
7200
O1: Adjust 20mA
Cal. Output 1
-----
Cal. Output 2
7200
O1: Adjust 20mA
Cal. Output 2
48
Calibrate Output 1
Note. When adjusting the 4 and 20mA output values, the display reading is unimportant
and is used only to indicate that the output is changing when the
pressed.
See below.
Cal. Output 2
Adjust 4mA
Set the milliammeter reading to 4mA.
Note. The analog output range selected in
affect the reading.
Refer to Fig. 6.11 for connection details.
Adjust 20mA
Set the milliammeter reading to 20mA.
Note. The analog output range selected in
affect the reading.
Refer to Fig. 6.11 for connection details.
Continued below.
Cal. Output 2
Operating Page – see Section 2.3.
Diss. Organics
Calibrate Output 2
Note. Output 2 calibration is identical to Output 1 calibration.
Continued on next page.
Option Board
Continued on next page.
Option Board
Operating Page – see Section 2.3.
Diss. Organics
and
(see Section 5.5) does not
Configure Outputs
(see Section 5.5) does not
Configure Outputs
keys are