4.
Turn the adjustment screw C to the desired angle
position.
The castor can be angled in 5 different positions
from 0° to 6°.
5.
Push the adjustment screw C back into position after
the adjustment.
6. Re-tighten the screws A and B.
The screws should be tightened with a torque of
20 +/- 2 Nm.
5 mm allen key
Castor height
Observe that only the height (89,4 mm) version of
the castor attachment is adjustable in height.
1.
Loosen the screw A with approximately 7 turns.
2. Loosen screw B with approximately 2 turns.
3.
Pull out the castor housing C and adjust it to a desired
height.
There are three possible fixed heights D.
1431580-C
A
A
B
4.
Push the castor housing C back into position after
adjustment.
5. Re-tighten the screws A and B.
The screws should be tightened with a torque of
20 +/- 2 Nm.
5 mm allen key
Adjusting the brakes
WARNING!
Reduced brake effect
– Incorrect adjustments of the brake can reduce
the effectiveness of the brake.
WARNING!
Risk of reduced brake effect
– The brakes must not be moved further forwards
or backwards than indicated by the warning
label D.
A
1. Check that the tires have the correct air pressure.
You can find the information about the air
pressure on the tire.
2. Loosen the screw A.
3. Slide the brake C forward or backwards to the desired
position.
4. Re-tighten the screw.
The distance between the brake shaft B and the
tire should be approximately 15 mm.
If further adjustments of the brakes are necessary, the brake
attachment can also be adjusted forwards or backwards:
A
B
C
C
B
Options