Safety Information - Canon CX-1 Operation Manual

Digital retinal camera
Hide thumbs Also See for CX-1:
Table of Contents

Advertisement

This instrument conforms to IEC 60601-1-2:2001.
For U.S.A. and Canada
• When the CX-1 is going to be operated at a voltage of 240V in USA or Canada, be sure to con-
nect the instrument to a center tapped voltage source.
• This instrument is a CLASS I EQUIPMENT according to UL 60601-1 and CAN/CSA C22.2
No. 601.1.
MEDICAL EQUIPMENT WITH RESPECT TO ELECTRICAL SHOCK,
FIRE AND MECHANICAL HAZARDS ONLY
IN ACCORDANCE WITH UL 60601-1 and CAN / CSA C22.2 No. 601.1
41C4
• Grounding reliability can only be achieved when the equipment is connected to an equivalent
receptacle marked "Hospital Only" or "Hospital Grade".
• The following mark indicates that the instrument is a Type B Applied Parts.
• The degree of protection against ingress of water is IPX0.
• This equipment is not suitable for use in the presence of a flammable anaesthetic mixture with air
or with oxygen or nitrous oxide.
• The mode of operation is Continuous Operation.
For EU Countries
• The following mark shows compliance of the instrument with Directive 93/42/EEC.
• This instrument is a CLASS I EQUIPMENT according to EN 60601-1.
• The following mark indicates that the instrument is a Type B Applied Parts according to EN
60601-1.
• The degree of protection against ingress of water is IPX0.
• This equipment is not suitable for use in the presence of a flammable anaesthetic mixture with air
or with oxygen or nitrous oxide.
• The mode of operation is Continuous Operation.
Für Deutschland
• Während des Betriebs liegt der Schalldruckpegel dieses Instruments bei 70 dB(A) oder weniger
gemäß ISO 7779.

Safety Information

Regulations
(1)