5.3.3 Cobalt Photography Mode

In this mode, photography using the exciter filter for fluorescein angiography is possible.
1
Enter the study information into the computer.
2
Select the cobalt photography mode.
Press the COBALT switch.
Or select the photography mode with the control
software.
The COBALT lamp turns on. COBALT is displayed on
the observation monitor.
3
Perform the procedure starting from step 3 in 5.3.1 Color Photography Mode (see page 14).

5.3.4 Fluorescein Angiography Mode

If taking multiple photographs, a mydriatic must be used.
1
Enter the study information into the computer.
2
Select the fluorescein angiography mode.
Press the FLUO switch.
Or select the photography mode with the control
software.
The FLUO lamp turns on. FLUO is displayed on the
observation monitor.
3
Disinfect the forehead rest and replace the chin
rest paper.
4
Have the patient sit down.
5
Align the height of the patient's eye.
6
Move the stage position.
7
Align the retinal camera to the eye to be photographed.
8
Adjust brightness in the field of view for observation.