User Calibration button

The
the sensor.
The
to indicate you have changed a setting and you have not
yet saved the setting to the Smart Manometer.
User Calibration: For authorized personnel only
What does "Calibration Enabled" refer to?
With a check mark in the box, Calibration Enabled displays the
Live Reading
Without a check mark in the box, it displays
without your user calibration.
Restore Factory Defaults
This button clears your user calibration from the sensor.
Choose Calibration Unit
Select the calibration reference units.
This does not affect or depend on the configuration units.
Begin button...Apply button
1.
Click Begin.
2. Select the point.
3. Apply reference pressure within the range shown.
4. Enter the reference value.
5. Save the point.
6. Repeat steps 2-5 until you are done. (You have to change at
least one point.)
7.
Click the
User Manual HHP460 Smart Manometer with Altimeter & Air Speed Test
User Calibration
button displays options to calibrate
User Calibration
button displays a
with your current user calibration applied.