Relief Valve Test - Omega HHP460 User Manual

Smart manometer with altimeter and air speed tester
Table of Contents

Advertisement

M20013 REL-0000 (2103)

Relief Valve Test

Starting the Relief Valve Test
To start the Relief Valve Test follow these steps:
1.
Press the
TAP START TO BEGIN
2.
MAX TEST P
maximum pressure measured during the test.)
3. Press the
minimum pressure measured after reaching MAX and pressure
starts releasing. If a new MAX is reached, MIN resets.)
4. Press the
Note: At this point, you see options available to you during the
Relief Valve Test.
1.
OR X TO EXIT TEST
2. 0
TO RESET MIN + MAX
3.
STP-FREEZE
briefly appears and the current value remains on-screen
blinking.
Press the
continue to display.
5. Press the
User Manual HHP460 Smart Manometer with Altimeter & Air Speed Test
Forward
key until you see
displays when the test begins. (MAX is the
Forward
key to see
MIN TEST P
Forward
key and you see
(use either option to exit)
(the
When you press the
Zero
key and
HOLD.OFF
Forward
key to see
MAX TEST P
RELIEF VALVE TEST
value. (MIN is the
--out-
.
Zero
key)
Stop
key,
hold.on
appears and new values
value.
33 of 72

Advertisement

Table of Contents
loading

Table of Contents