1. Remove right power module shroud and rear shroud.
2. Unplug battery cable on power module.
3. Disconnect circuit breaker cables from batteries.
Disconnect negative terminal first, then positive
terminal.
a. Remove tie wraps (not shown).
b. Remove terminal caps A.
c. Loosen and remove terminal screws B together with
washers C and cable lug from terminal.
4. Loosen clip D.
5. Remove circuit breaker together with cables.
Replacing Holder
7 mm wrench
1654647-B
Fig. 6-54 Positive terminal serves as an example.