High Voltage Probe HV-P60A

4.2 Measuring Procedure

(1) Confirm that the voltage of the object to be measured is within
the range of the measurement of this instrument.
(2) Confirm that there is neither person in surroundings nor
something that touches.
(3) The power supply of the object to be measured is turned on,
and the measurement begins.
Note
When measuring, ringing and the overshoot might be
caused. In addition, it is necessary to strengthen GND in
addition to the earth lead of the attachment to decrease ringing
and the overshoot. Prepare the metal with high electric
conductivity (treatment device), tie to the earth lead joint, and
strengthen GND. In that case, do not touch the probe point with
GND.
As a concrete example, there are the following methods of etc.
(1) The stick made of copper (about φ5 mm) is used.
(2) The thin sheet copper is folded and used.
Example) The thin sheet copper plate of the width of 50mm is
* However, use together or substitute the earth lead of the
attachment if necessary when you use the above-mentioned
treatment device.
folded four times, and it is used by the width of 10mm.
12