3.2 Adjustment

This instrument is connected and adjusted according to the following
procedures.
(1) As shown in Figure 3(a), the ground terminal of the probe body is
connected with the terminal GND of the signal generator.
(2) INPUT of the oscilloscope is connected with the connector of the
probe body with Output cable.
(3) The output of the signal generator is connected with the tip of
the probe body
Connector
Oscilloscope
Matching Box
Figure 3 (a) Connection when probe is adjusted
Probe body
Ground terminal
Ground lead
Hole for waveform adjustment
Figure 3
(b) Waveform adjustment of probe
in matching box (closeup)
High Voltage Probe HV-P60A
Tip of Probe
Clamp
Screwdriver
5
Signal Generator